Output switching (:ext:outp) – GW Instek PSB-2000 Series User Manual User Manual
Page 121
EXTERNAL CONTROL
121
External Control of the Current Output ON/OFF
(:EXT:CURR)
This command chooses whether the external control is used to
control the output current or if the front panel controls are used to
control the output current.
Setting
:EXT:CURR
Activates or deactivates external control of the output
current of a single channel model.
:EXT:CURR:A
Activates or deactivates external control of the output
current of CH1.
:EXT:CURR:B
Activates or deactivates external control of the output
current of CH2.
0: Deactivates external control of the output current
(Allows the current output to be controlled with the front
panel controls or via remote control).
1: Activates external control of the current output.
Application
example
:EXT:CURR 1
Selects the external control of the current output.
Query
:EXT:CURR?
Inquires about the setting.
Response
example
1
Indicates that external control of the current output is
enabled. The query return value has the same meaning as
the setting
Remark
This command causes an error during sequence
operation (RUN).
Output switching (:EXT:OUTP)
This command selects output On/Off with external contacts.
Setting
:EXT:OUTP
0: Control from the panel or through communication
1: Output On/Off with external contacts
Application
example
:EXT:OUTP 1
Selects output On/Off with external contacts.
Query
:EXT:OUTP?
Inquires about the setting.