5.

D

ATA

P

ATTERN

S

ETUP

Data Value - Provide an appropriate value based on your Binary or Hexadecimal format selection.

Data Value To - Specify a size in bits for your pattern.

Condition - Possible entries for this field include Equal, Not Equal, Less than, Less than or Equal to,
Greather than, Greater than or Equal to, In Range, or Out Range.

Pattern Length - The pattern length value defaults to the length, in bytes, of the pattern set in the Data
Value selection. If the length is changed to a lesser value, it truncates the beginning of the value. If the
length is increased, it would add don’t care XX byte values to the beginning of the value.

6.

I

NTER

F

RAME

S

ETUP

Click the appropriate button to select None or Manual. The Manual button enables the Byte Position and
Interframe Time