62
Acyclic Data Transfer with
PROFIBUS, DeviceNet and CANopen
Bürkert mfc family Object
name
Description of the input
data attributes
attribute address
(class, instance
attribute; data type)
Slot, index
index, Subindex
Dvn
Dpv1
canopen
Gas1
SafeValue
Read Write
Change in the flow rate of
gas1 to which the device is
set in case of an emergency
in the system
(value in units per thousand)
min. value 0, max. value 1000
dec: 110, 1, 30
Hex: 6E, 1, 1E
UINT
dec: 1, 180
Hex: 1, B4
dec: 12288, 30
Hex: 3000, 1
EUNSIGNEd16
Gas2
SafeValue
Read Write
Change in the flow rate of
gas2 to which the device is
set in case of an emergency
in the system
(value in units per thousand)
min. value 0, max. value 1000
dec: 110, 1, 31
Hex: 6E, 1, 1F
UINT
dec: 1, 181
Hex: 1, B5
dec: 12288, 31
Hex: 3000, 1F
UNSIGNEd16
Binary
output 1
function-limits
Read Write
determines when binary
output 1 is active. It is a
logical OR operation of all
"Binary output 1 functions"
(in this case: group of limits;
see “Bit field LIMITS” in
“9.1. description of bit fields”
min. value 0,
max. value 65535
dec: 110, 1, 32
Hex: 6E, 1, 20
WORd
dec: 1, 182
Hex: 1, B6
dec: 12288, 32
Hex: 3000, 20
UNSIGNEd16
Binary
output 1
function-
errors
Read Write
determines when binary
output 1 is active. It is a
logical OR operation of all
"Binary output 1 functions"
(in this case: group of errors;
see “Bit field ERRORS” in
“9.1. description of bit fields”
min. value 0,
max. value 65535
dec: 110, 1, 33
Hex: 6E, 1, 21
WORd
dec: 1, 183
Hex: 1, B7
dec: 12288, 33
Hex: 3000, 21
UNSIGNEd16
Binary
output 1
function-
others
Read Write
determines when binary
output 1 is active. It is a
logical OR operation of all
"Binary output 1 functions"
(in this case: group of others;
see “Bit field OTHERS”
in “9.1. description of bit
fields”) min. value 0,
max. value 65535
dec: 110, 1, 34
Hex: 6E, 1, 22
WORd
dec: 1, 184
Hex: 1, B8
dec: 12288, 34
Hex: 3000, 22
UNSIGNEd16
english
MFC Family