INFICON IC/5 Thin Film Deposition Controller User Manual
Page 83
3 - 31
IP
N 07
4-
23
7A
E
IC/5 Operating Manual
12. RATE RAMP TIME 2
Rate control, desired rate
changing. [New Rate 2, Start Ramp
2, Ramp Time 2]
Active
Active
13. RATEWATCHER
(SAMPLE)
Rate control. [RateWatch
Accuracy]
Active
Active
14. RATEWATCHER
(HOLD)
Constant power, based on last
sample’s average power.
[RateWatch Time]
Active
Inactive
15. MANUAL
Source power controlled by
hand-held controller.
Active
Active
16. TIME-POWER
Crystal failed; source maintained at
average control power prior to
crystal failure. [Time Pwr Y]
Active
Active
NOTE: 17 through 20 are Post-Deposit states.
17. FEED RAMP
Source changing to Feed Power.
[Feed Power, Feed Ramp Time]
Inactive
Inactive
18. FEED
Source maintained at Feed Power
level. [Feed Time]
Inactive
Inactive
19. IDLE RAMP
Source changing to Idle Power.
[Idle Ramp Time, Idle Power]
Inactive
Inactive
20a.IDLE POWER (=0%)
Source maintained at zero power;
will accept a START command.
Inactive
Inactive
20b.IDLE POWER (>0%)
Source resting at Idle Power; will
accept a START command.
Inactive
Inactive
21. STOP
Source output set to zero power.
The display is frozen at the last rate
and thickness values; will not
accept a START or RESET
command.
Inactive
Inactive
NOTE: In the STOP state the instrument will accept a START
provided a Crystal Failure has not occurred for the sensors
used in the layer being started.
Table 3-3 State Descriptions (continued)
STATE
CONDITION
RELAY CONTACT
STATUS
NOTE: 1 through 7 are Pre-Deposit states.
Source
Shutter
Sensor
Shutter