4 update commands, 1 update material parameters – INFICON IC/5 Thin Film Deposition Controller User Manual
Page 160
7 - 18
IP
N 07
4-
23
7A
E
IC/5 Operating Manual
7.7.4 UPDATE Commands
Update commands are used to change a specified parameter to a new value or
condition. Some parameters are order dependent in that they cannot be updated
prior to certain other parameters. For example, in the IC/5 two layers must be
defined before specifying the codeposition of those layers.
There is a specific update command for each parameter group. Each command
has one to four arguments starting with and separated by a space. The Update
command variations include:
7.7.4.1 Update Material Parameters
For each command format the symbol
þ
represents a blank space.
The format is:
UM
þ
param#
þ
material#
þ
value
This command changes the value of a material parameter for a specified material.
Refer to
, for parameter value ranges.
Table 7-11 Update Commands
command
parameter
UM
Update Material Parameters
UP
Update Process Parameters
UU
Update Utility Parameters
US
Update Sensor Parameters
UC
Update Source Control Parameters
UT
Update Output Relay Type
UL
Update Logic Statement
UO
Update alphanumeric name of hardware output
UI
Update alphanumeric name of hardware input
UN
Update user messages
Table 7-12 Update Material Parameters
param# Material
Parameter
1
Density
2
Z-Ratio
3
Source
4
Control Loop
5
Process Gain
6
Primary Time Constant
7
System Dead Time
8
Recorder Output
9
Recorder Function
10
Master Tooling