INFICON IC/5 Thin Film Deposition Controller User Manual
Page 204
9 - 6
IP
N 07
4-
23
7A
E
IC/5 Operating Manual
FLOPPY DISK ACCESS CODE . . . 1 to 9999, 0
The Floppy Disk Access Code allows configuration file retrieval from a floppy
disk when the Program Lock Code is locked. This allows a process recipe to be
entered into the IC/5 from a floppy disk without “unlocking” the Program Lock
Code. Otherwise floppy disk retrieval is not allowed when a Program Lock
Code is present. This parameter has no effect if there are no lock codes, or the
access code is set to zero.
NOTE: When pressing the Clear key on power-up, the Floppy Disk Access
Code is cleared when the parameters are cleared; not when the lock
codes are cleared. The Floppy Disk Access Code is not stored on nor
retrieved from a floppy disk. Program and I/O lock codes are stored on
and retrieved from a floppy disk.
SYSTEM TIME. . . . . . . . . . . . . . . . . 00:00 to 23:59 mins:secs
Twenty-four-hour clock for current time of day. Values range from 00:00 to
23:59.
SYSTEM DATE . . . . . . . . . . . . . . . . mm-dd-yyyy
Current date. Increments automatically. Values range from 01-01-0000 to
12-31-9999.
AUDIO FEEDBACK. . . . . . . . . . . . . YES/NO
Activates Audio Feedback; signifying data entry. The default value is Yes which
activates the audio signal.
CODEP DISPLAY . . . . . . . . . . . . . . 0/1
This value determines whether the Operate Display for a co-deposition
provides text-only or a text/graphic display. The default is 0 for a text/graphic
display.
RATE SMOOTHING. . . . . . . . . . . . . YES/NO
This parameter provides an averaging of the instantaneous rate measurement.
The default value is YES. When set to YES, the Rate displayed in the graphics
area of the OPERATE display and the rate output to the analog output is a 5
point box car average. When set to NO, the rate displayed in the graphics area
of the OPERATE display and the rate output to the analog output is the
instantaneous rate measurement. The measurement rate is 10 Hz.