Layer_status_code, Layer, Global_status_code – INFICON IC/5 Thin Film Deposition Controller User Manual
Page 175
7 - 33
IP
N 07
4-
23
7A
E
IC/5 Operating Manual
SL
þ
layer_status_code
þ
Layer #
These commands are used to determine the condition of started layers. For
example: SL 1 1 will return the current Rate for Layer 1.
For the Response Formats: x represents a number (if the value being returned
does not occupy the entire width, blanks are added to the left); s represents the
sign (only a negative sign is displayed, a blank is returned in this position if the
value is positive); t represents the timer tick value; w represents the sensor status
byte and
þ
represents a blank.
11
Defined Material Count
The value returned indicates the number of
materials that are defined.
xx
12
Timer Tick and the delta fundamental frequency for
all eight sensors. 1-9 returned for a positive change.
A-I returned for a negative change. 0 returned for no
change.
ttttxxxxxxxx
Table 7-20 Status Command
SL
þ
layer_status_code
þ
Layer #
Code
Channel Condition
Response Format
0
Process information. All the information for codes 1
to 9
1
Rate (Å/s) currently read. (Raw reading) sx.xxxb to
sxxxx.b
sx.xxx
þ
to sxxxx.
þ
2
Power (%) currently read.
þ
xx.xxx
þ
3
Thickness (KÅ) currently accumulated
sxxx.xxx
þ
4
State currently in progress.
þ
xx
þ
Response Codes to SL 04 Layer # Command.
State#
State in Process
0
READY
1
SOURCE SWITCH
2
RISE 1
3
SOAK 1
4
RISE 2
5
SOAK 2
6
SHUTTER DELAY
7
DEPOSIT
(before any rate ramps)
Table 7-19 Status Command
SG
þ
global_status_code
Code#
Global Condition
Response Format