7.7.4.8 Update Output Name
For each command format the symbol
þ
represents a blank space.
The format is:
UO
þ
output#
þ
output string
This command will assign a user-defined string to the output # designated unless
the output already has an IC/5-assigned name. The string can be no more than 10
characters long and must be comprised from the character set:
__ABCDEFGHIJKLMNOPQRSTUVWXYZ0123456789!#~-%&
The output string may be cleared by sending the UO command with the output
string left blank. The underscore is used to represent a blank space.
7.7.4.9 Update Input Name
For each command format the symbol
þ
represents a blank space.
The format is:
UI
þ
input#
þ
input string
This command will assign a user-defined string to the input # designated unless the
input already has an IC/5-assigned name. The string can be no more than 10
characters long and must be comprised from the character set:
__ABCDEFGHIJKLMNOPQRSTUVWXYZ0123456789!#~-%&
The input string may be cleared by sending the UI command with the input string
left blank. The underscore is used to represent a blank space.
7.7.4.10 Update User Message
For each command format the symbol
þ
represents a blank space.
The format is:
UN
þ
user_message#
þ
user message string
This command will assign a user-defined string to the user message number
designated. The string can be no more than 19 characters in length and must be
comprised from the character set:
__ABCDEFGHIJKLMNOPQRSTUVWXYZ0123456789!#~-%&
The user message string may be cleared by sending the UN command with the
user message string left blank. The underscore is used to represent a blank space.