Model GFC7000E Instruction Manual

Operating Instructions

04584 Rev A1

81

6.9.9. Test Channel Output

When activated, output channel A3 can be used to report one of the test functions viewable from

the SAMPLE mode display. To activate the A3 channel and select a test function, follow this key
sequence:

SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

DIAG

SIGNAL I / O

NEXT

ENTR

EXIT

DIAG

ANALOG OUTPUT

PREV

NEXT

ENTR

EXIT

DIAG TCHN

TEST CHANNEL: NONE

NEXT

ENTR

EXIT

DIAG

TEST CHAN OUTPUT

PREV

NEXT

ENTR

EXIT

EXIT returns

to the main

SAMPLE

display

Continue to press NEXT until …

DIAG TCHN

TEST CHANNEL: CO2 MEASURE

PREV

NEXT

ENTR

EXIT

Press PREV or NEXT

to move through the

list of available

parameters

(Table 6-13)

Press EXIT to

return to the

DIAG menu

Press ENTR to select

the displayed

parameter activating

the test channel.

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

ALRM

EXIT

Table 6-14: Test Parameters Available for Analog Output A3

Test Channel

Zero

Full Scale

NONE

Test Channel is turned off

CO

2

MEASURE

0 mV

5000 mV*

CO

2

REFERENCE

0 mV

5000 mV*

SAMPLE PRESS

0 "Hg

40 "Hg

SAMPLE FLOW

0 cc/m

1000 cc/m

SAMPLE TEMP

0°C 70°C

BENCH TEMP

0°C 70°C

WHEEL TEMP

0°C 70°C

CHASSIS TEMP

0°C 70°C

PHT DRIVE

0 mV

5000 mV

* This refers to the internal voltage level of the function NOT the output

signal level of the Test channel itself.