Teledyne GFC-7000E - Trace CO2 Analyzer User Manual
Page 253
Model 360E Instruction Manual
APPENDIX A-4: MGFC7000E Signal I/O Definitions, Revision E.0
045840110 Rev B.3
242
SIGNAL NAME
BIT OR CHANNEL
NUMBER
DESCRIPTION
CO_MEASURE
2
Detector measure reading
CO_REFERENCE
3
Detector reference reading
4
Temperature
MUX
SAMPLE_FLOW 5
Sample
flow
PHOTO_TEMP
6
Photometer detector temperature
TEST_INPUT_7
7
Diagnostic test input
TEST_INPUT_8
8
Diagnostic test input
REF_4096_MV
9
4.096V reference from MAX6241
O2_SENSOR
5
10
O
2
concentration sensor
11
Spare
CO2_SENSOR
7
12
CO
2
concentration sensor
13
Spare
14
DAC
loopback
MUX
REF_GND 15
Ground
reference
Rear board temperature MUX analog input
BOX_TEMP 0
Internal
box
temperature
SAMPLE_TEMP 1
Sample
temperature
BENCH_TEMP
2
Optical bench temperature
WHEEL_TEMP 3
Wheel
temperature
TEMP_INPUT_4
4
Diagnostic temperature input
TEMP_INPUT_5
5
Diagnostic temperature input
O2_CELL_TEMP 6
O
2
sensor cell temperature
7
Spare
Rear board DAC MUX analog inputs
DAC_CHAN_0
0
DAC channel 0 loopback
DAC_CHAN_1
1
DAC channel 1 loopback
DAC_CHAN_2
2
DAC channel 2 loopback
DAC_CHAN_3
3
DAC channel 3 loopback
Rear board analog outputs
CONC_OUT_1
0
Concentration output #1
CONC_OUT_2
1
Concentration output #2
TEST_OUTPUT
2
Test measurement output
CONC_OUT_3
7
3
Concentration output #3 (CO
2
)
1
Hessen
protocol.
2
M300EH.
3
M300ES.
4
M320E.
5
O
2
option.
6
Sample pressure or differential pressure flow measurement option.
7
M306E.
8
Factory option enables concentration alarms on relay outputs.
9
GFC7000E.
10
GFC7000E.