Model GFC7000E Instruction Manual

Calibration Procedures

04584 Rev A1

140

The following example sets Sequence #2 to do a Zero-Span Calibration every other day starting
at 1 am on September 4, 2001, lasting 15 minutes, without calibration. This will start ½ hour later

each iteration.

MODE AND

ATTRIBUTE

VALUE COMMENT

Sequence 2

Define Sequence #2

Mode ZERO-SPAN

Select Zero and Span Mode

Timer Enable

ON

Enable the timer

Starting Date

Sept. 4, 2001

Start after Sept 4, 2001

Starting Time

01:00

First Span starts at 1:00AM

Delta Days

2

Do Sequence #2 every other day

Delta Time

00:30

Do Sequence #2 ½ hr later each day

Duration 15.0

Operate Span valve for 15 min

Calibrate NO

Do not calibrate at end of Sequence

NOTE

The programmed STARTING_TIME must be a minimum of 5 minutes later than the real

time clock (See Section 6.6 for setting real time clock).

NOTE

Avoid setting two or more sequences at the same time of the day. Any new sequence

which is initiated whether from a timer, the COM ports, or the contact closure inputs

will override any sequence which is in progress.

NOTE

If at any time an illegal entry is selected (Example: Delta Days > 367) the ENTR key will

disappear from the display.