Model GFC7000E Instruction Manual

TROUBLESHOOTING & REPAIR PROCEDURES

04584 Rev A1

208

please use the following procedure (see the Spare Parts list in Appendix B for part numbers and
kits):

1. Turn off power to the analyzer.

2. Locate the assembly attached to the sample pump, see Figure 3–11.

3. Disconnect the pneumatic connection from the flow assembly and the assembly from the

pump.

4. Remove the fitting and the components as shown in the exploded view in Figure 11.6.

5. Replace the o-rings (p/n:OR_01) and the sintered filter (p/n:FL_01).

6. If replacing the critical flow orifice itself (p/n:00094100), make sure that the side with the

colored window (usually red) is facing upstream to the flow gas flow.

7. Re-assemble in reverse order.

8. After reconnecting the power and pneumatic lines, flow check the instrument as described in

the Section 9.3.4.

Pneumatic Connector, Male 1/8”
(P/N FT_70

Spring
(P/N HW_20)

Sintered Filter
(P/N FL_01)

Critical Flow Orifice
(P/N 00094100)

O-Ring

(P/N OR_01)

Purge Housing

(P/N 000850000)

Figure 11-6: Critical Flow Restrictor Assembly Disassembly