TROUBLESHOOTING & REPAIR PROCEDURES

04584 Rev A1

185

Table 11-1: Warning Messages - Indicated Failures

WARNING

MESSAGE

FAULT CONDITION

POSSIBLE CAUSES

BENCH TEMP

WARNING

The optical bench temp

is controlled at 48 ± 5

°

C.

Bad bench heater
Bad bench temperature sensor

Bad relay controlling the bench heater

Entire relay board is malfunctioning

I

2

C

buss malfunction

BOX TEMP

WARNING

Box Temp is < 5

°

C or

> 48

°

C.

NOTE: Box temperature typically runs ~7

o

c warmer than

ambient temperature.

Poor/blocked ventilation to the analyzer.
Stopped exhaust-fan

Ambient temperature outside of specified range

CANNOT DYN

SPAN

Dynamic Span

operation failed

Measured concentration value is too high or low.

Concentration slope value to high or too low

CANNOT DYN

ZERO

Dynamic Zero

operation failed

Measured concentration value is too high.

Concentration offset value to high.

CONFIG

INITIALIZED

Configuration and

Calibration data reset

to original Factory

state.

Failed disk on chip

User erased data

CONC ALRM1

WARNING

Concentration alarm 1

is enabled and the
measured CO

2

level is

≥ the set point.

CONC ALRM2

WARNING

Concentration alarm 2

is enabled and the

measured CO

2

level is

≥ the set point.

DATA

INITIALIZED

Data Storage in iDAS

was erased

Failed disk on chip

User cleared data

FRONT PANEL

WARN

The CPU is unable to

Communicate with the
Front Panel Display

/Keyboard

Warning only appears on serial I/O com port(s)

Front panel display will be frozen, blank or will not respond.
Failed keyboard

I

2

C

buss failure

Loose connector/wiring

PHOTO TEMP

WARNING

PHT DRIVE is >2500
mVDC

Failed IR photo-detector
Failed sync/demod board

IR photo-detector improperly attached to the sample

chamber
Bench temp too high.

REAR BOARD

NOT DET

Mother Board not
detected on power up.

Warning only appears on serial i/o com port(s)
Front panel display will be frozen, blank or will not respond.

Massive failure of mother board

RELAY BOARD

WARN

The CPU cannot
communicate with the

Relay Board.

I

2

C

buss failure

Failed relay board
Loose connectors/wiring

SAMPLE FLOW

WARN

Sample flow rate is <
500 cc/min or > 1000

cc/min.

Failed sample pump
Blocked sample inlet/gas line

Dirty particulate filter

Leak downstream of critical flow orifice

Failed flow sensor/circuitry