Model GFC7000E Instruction Manual

Operating Instructions

04584 Rev A1

47

Title

Range

RANGE

RANGE1

1

RANGE2

1

PPB, PPM,

UGM,

MGM

The full scale limit at which the reporting range of the
analyzer is currently set.
THIS IS NOT the Physical Range of the instrument. See
Section 6.7 for more information.

Stability

STABIL

PPB, PPM

UGM,

MGM

Standard deviation of CO

2

concentration readings. Data

points are recorded every ten second. The calculation uses
the last 25 data points.

CO

2

Measure

CO

2

MEAS

MV

The demodulated, peak IR detector output during the
measure portion of the CFG Wheel cycle.

CO

2

Reference

CO

2

REF

MV

The demodulated, peak IR detector output during the

reference portion of the CFG wheel cycle.

Measurement /
Reference
Ratio

MR Ratio

-

The result of CO2 MEAS divided by CO2 REF. This ratio is
the primary value used to compute CO

2

concentration. The

value displayed is not linearized.

Sample
Pressure

PRES

In-Hg-A

The absolute pressure of the Sample gas as measured by a

solid state pressure sensor located inside the sample
chamber.

Sample Flow

SAMPLE

FL

cc/min

Sample mass flow rate. This is computed from the differential

between the pressures measured up-stream and down-
stream of the sample critical flow orifice pressures.

Sample

Temperature

SAMP

TEMP

°C

The temperature of the gas inside the sample chamber.

Bench
Temperature

BENCH

TEMP

°C

Optical bench temperature.

Wheel
Temperature

WHEEL

TEMP

°C

Filter wheel temperature.

Box
Temperature

BOX

TEMP

°C

The temperature inside the analyzer chassis.

Photo-detector
Temp. Control

Voltage

PHT

DRIVE

mV

The drive voltage being supplied to the thermoelectric coolers
of the IR photo-detector by the sync/demod Board.

Slope

SLOPE

-

The sensitivity of the instrument as calculated during the last

calibration activity. The SLOPE parameter is used to set the
span calibration point of the analyzer.

Offset

OFFSET

-

The overall offset of the instrument as calculated during the

last calibration activity. The OFFSET parameter is used to
set the zero point of the analyzer response.

Test channel
output signal

TEST

mV, mA

Displays the signal level of the TEST analog output channel.
Only appears when the TEST channel has been activated.

Current Time

TIME

-

The current time. This is used to create a time stamp on

iDAS readings, and by the AUTOCAL feature to trigger
calibration events.

1

Only appears when the instrument’s reporting range mode is set for DUAL or AUTO