Model, Span gas concentration – Teledyne GFC-7000E - Trace CO2 Analyzer User Manual
Page 146
Model GFC7000E Instruction Manual
Calibration Procedures
04584 Rev A1
135
Figure 7-5: Pneumatic Connections–MGFC7000E with Zero/Span Valves (OPT 52)
External Zero
Air Scrubber
Certified
CO Gas
Source of
SAMPLE Gas
VENT if input is pressurized
MODEL 700
Gas Dilution
Calibrator
VENT
MODEL 701
Zero Air Generator
Indicating
soda-lime
MODEL
GFC7000E
Sample
IZS
Exhaust
Vent Span
Pressure Span
Purge In
Figure 7-6: Pneumatic Connections–MGFC7000E with Zero/Span Valves with External
Zero air Scrubber (OPT 53)
STEP TWO: Set the expected CO
2
Span Gas concentration:
This sequence causes the
analyzer to prompt for the
expected CO
x
span
concentration.
M-P CAL
SO2 SPAN CONC: 450.000 Conc
0 0 0
4 5 .0
ENTR
EXIT
M-P CAL RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > ZERO
CONC
EXIT
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST >
CAL
CALZ CALS
SETUP
The SO
2
span
concentration values
automatically default to
400.0 Conc.
To change this value to
the actual concentration of
the span gas, enter the
number by pressing the
key under each digit until
the expected value
appears.
EXIT ignores the new setting
and returns to the previous
display.
ENTR accepts the new setting
and returns to the
previous display..
NOTE
For this Initial Calibration it is important to independently verify the PRECISE CO
2
Concentration Value of the SPAN gas.
If the source of the Span Gas is from a Calibrated Bottle, use the exact concentration
value printed on the bottle.