beautypg.com

Teledyne GFC-7000E - Trace CO2 Analyzer User Manual

Page 25

background image

Model GFC7000E Instruction Manual

Getting Started

04584 Rev A1

14

The pin assignments for the status outputs can be found in the table below:

Table 3-2: GFC7000E Status Output Pin Outs

Output #

Status

Definition

Condition

1

SYSTEM OK

On if no faults are present.

2

CONC VALID

On if CO

2

concentration measurement is valid.

If the CO

2

concentration measurement is invalid, this bit is OFF.

3

HIGH RANGE

On if unit is in high range of DUAL or AUTO range modes.

4

ZERO CAL

On whenever the instruments ZERO point is being calibrated.

5

SPAN CAL

On whenever the instruments SPAN point is being calibrated.

6

DIAG MODE

On whenever the instrument is in DIAGNOSTIC mode.

7

ALARM1

On whenever the measured CO

2

concentration is above the set

point for ALM1

8 ALARM2

On whenever the measured CO

2

concentration is above the set

point for ALM2

D

EMITTER BUSS The emitters of the transistors on pins 1-8 are bussed together.

+

DC POWER

+ 5 VDC

Digital Ground

The ground level from the analyzer’s internal DC power supplies.

11. If you wish to use the analyzer’s to remotely activate the zero and span calibration modes,

several digital control inputs are provided via a 10-pin connector labeled CONTROL IN on the

analyzer’s rear panel. Two methods for energizing the inputs are provided below; the first
using the internal +5V available on the CONTROL IN connector and the second, if an external,

isolated supply is employed.

CONTROL IN

A B C D E F U +

S
P

A

N

Z
E

R
O

CONTROL IN

A B C D E F U +

-

+

5 VDC Power

Supply

S

P
A

N

Z

E

R

O

Local Power Connections

External Power Connections