Communication options, Zero/span valve with external co, Scrubber (option 53) – Teledyne GFC-7000E - Trace CO2 Analyzer User Manual

Page 52: Sample chamber

background image

Model GFC7000E Instruction Manual

Optional Hardware and Software

04584 Rev A1

41

INSTRUMENT CHASSIS

PUMP

SAMPLE CHAMBER

VENT SPAN

OUTLET

PRESSURE

SPAN INLET

IZS INLET

EXHAUST GAS

OUTLET

PURGE GAS

INLET

SAMPLE GAS

INLET

FLOW

SENSOR

SAMPLE

PRESSURE

SENSOR

FLOW / PRESSURE
SENSOR PCA

S

a

m

p

le

G

a

s

C

rit

ic

a

l

F

low

Or

if

ic

e

GFC Wheel

Motor

GFC Motor

Heat Sync

GFC Wheel

Housing

Purge Gas

Pressure

Control Assy

PARTICULATE

FILTER

SAMPLE /CAL

VALVE

3

2
1

Purge Gas

Flow Control

Orifice

ZERO/SPAN
VALVE

2
1
3

Ex

te

rn

a

l C

O

2

Sc

ru

b

b

er

( Op

ti

o

n

5

3

On

ly

)

Figure 5-3: Internal Pneumatic Flow – Zero/Span OPT 52 & 53

5.4.4. Zero/Span Valve with External CO

2

Scrubber (Option 53)

Option 53 is operationally and pneumatically identical to Option 52 above (see Section 5.4.3),
except that the zero air is generated by an externally mounted zero air scrubber filled with

indicating soda-lime that changes color from white to pink as it becomes saturated.

5.5. Communication Options

5.5.1. RS232 Modem Cable (Option 60)

This option consists of a cable to connect the analyzer’s COM1 port to a computer, a code

activated switch or any other communications device that is equipped with a DB-9 male
connector. The cable is terminated with two DB-9 female connectors, one of which fits the
analyzer’s COM1 port.

Some older computers or code activated switches with a DB-25 serial connector will need a
different cable or an appropriate adapter.