6.11. SETUP – ALRM: Using the Gas Concentration

Alarms

The Model GFC7000E includes two CO

2

concentration alarms Each alarm has a user settable

limit, and is associated with an opto-isolated TTL relay accessible via the status output connector
on the instrument’s back panel (see Section 6.13.1.1). If the CO

2

concentration measured by the

instrument rises above that limit, the alarm‘s status output relay is closed.

The default settings for ALM1 and ALM2 are:

Table 6-20: CO

2

Concentration Alarm Default Settings

ALARM STATUS

LIMIT SET POINT

1

ALM1

Disabled 100

ppm

ALM2

Disabled 300

ppm

1

Set points listed are for PPM. Should the reporting range units of measure be changed (see Section

6.7.6) the analyzer will automatically scale the set points to match the new range unit setting.

Note

To prevent the concentration alarms from activating during span calibration operations

make sure to press CAL or CALS button prior to introducing span gas into the analyzer.

To enable either of the CO

2

concentration alarms and set the Limit points, press:

ENTR key accepts the

new settings

EXIT key ignores the

new settings

SETUP X.

ALARM 1 LIMIT: 200,00 PPM

0 1 0 0 .0 0

ENTR EXIT

SETUP X.X

ALARM MENU

ALM1 ALM2

EXIT

SETUP X.

ALARM 1 LIMIT: OFF

OFF

ENTR EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS DIAG

ALRM

EXIT

SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

Toggle these keys to

cycle through the

available character set:

0-9

SETUP X.

ALARM 1 LIMIT: ON

ON

ENTR EXIT