beautypg.com

Model, Gfc7000e – Teledyne GFC-7000E - Trace CO2 Analyzer User Manual

Page 27

background image

Model GFC7000E Instruction Manual

Getting Started

04584 Rev A1

16

CAUTION

In order to prevent dust from getting into the gas flow channels of your analyzer, it was

shipped with small plugs inserted into each of the pneumatic fittings on the back panel.

Make sure that all of these dust plugs are removed before attaching

exhaust and supply gas lines.

Table 3-4: Model GFC7000E Rear Panel Pneumatic Connections

Rear Panel Label

Function

SAMPLE

Connect a gas line from the source of sample gas here.
Calibration gasses are also inlet here on units without
zero/span/shutoff valve or IZS options installed.

EXHAUST

Connect an exhaust gas line of not more than 10 meters long here.

PRESSURE SPAN

On units with zero/span/shutoff valve options installed, connect a
gas line to the source of calibrated span gas here.

VENT SPAN

Span gas vent outlet for units with zero/span/shutoff valve options
installed.
Connect an exhaust gas line of not more than 10 meters long here.

IZS

Internal zero air scrubber.

on units with zero/span/shutoff valve options installed but NO
internal zero air scrubber, attach a gas line to the source of zero air

here.

PURGE IN

This inlet supplies purge air to the GFC wheel housing (see section
10.2.2)
Connect a source of dried air that has been scrubbed of CO

2

.

VENT

Calibrated CO

2

gas at desired

span gas

concentration

Source of

SAMPLE Gas

Removed

during

Calibration

Valve

Needle

valve to

control

flow

MODEL

GFC7000E

Sample

IZS

Exhaust

Vent Span

Pressure Span

Purge In

Indicating

soda-lime

MODEL 701

Zero Air

Generator

Figure 3-3: Pneumatic Connections–Basic Configuration–Using Bottled Span Gas