Teledyne GFC-7000E - Trace CO2 Analyzer User Manual
Page 243
Model GFC7000E Instruction Manual
APPENDIX A-2: Setup Variables For Serial I/O, Revision
E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0
045840110 Rev B.3
232
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT VALUE
VALUE RANGE
DESCRIPTION
USER_UNITS
— PPM
0
PPB,
PPM,
3, 8
,
UGM, MGM
3, 8
,
Concentration units for user
interface.
DIL_FACTOR
—
1
0.1–1000
Dilution factor. Used only if is
dilution enabled with
FACTORY_OPT variable.
DARK_CAL_DURATION
Seconds 180,
60
4
10–600
Duration of dark cal. First
two-thirds is stabilization
period; final third is measure
period.
DARK_MEAS_MV
mV
0
-1000–1000
Dark offset for measure
reading.
DARK_REF_MV
mV
0
-1000–1000
Dark offset for reference
reading.
LIN_TARGET_CONC1
Conc
300
1–10000
Target concentration during
linearity adjustment for range
1.
LIN_NORM_CONC1
PPM
300
0.01–10000
Target concentration during
linearity adjustment
normalized for T/P for range
1.
LIN_RATIO1
— 1
0.01–100
Measure/reference
ratio
measured during linearity
adjustment for range 1.
LIN_CORRECT1
—
1
0.001–999.999
Linearity correction factor for
range 1.
LIN_TARGET_CONC2
Conc
300
1–10000
Target concentration during
linearity adjustment for range
2.
LIN_NORM_CONC2
PPM
300
0.01–10000
Target concentration during
linearity adjustment
normalized for T/P for range
2.
LIN_RATIO2
— 1
0.01–100
Measure/reference
ratio
measured during linearity
adjustment for range 2.
LIN_CORRECT2
—
1
0.001–999.999
Linearity correction factor for
range 2.
LIN_TARGET_CONC
Conc
300
1–10000
Target concentration during
linearity adjustment.
LIN_NORM_CONC
PPM
300
0.01–10000
Target concentration during
linearity adjustment
normalized for T/P.
LIN_RATIO
— 1
0.01–100
Measure/reference
ratio
measured during linearity
adjustment.
LIN_CORRECT
—
1
0.001–999.999
Linearity correction factor.
CO2_COMP_ENABLE
—
OFF
ON, OFF
ON enables CO
2
compensation; OFF disables
it.
CO2_COMP_CONC
% 0
0–20
CO
2
concentration to
compensate for.
CO_CONST1
— 700,
500
4
, 40000
3, 8
100–50000
CO
calculation
constant.
CO_CONST2
— 0.13,
0.198
8
1.448
4
,1.458
12
, 0.187
3
0–10 CO
calculation
constant.
ET_MEAS_GAIN
—
1
0.0001–9.9999
Electrical test gain factor for
measure reading.
ET_REF_GAIN
—
1
0.0001–9.9999
Electrical test gain factor for
reference reading.
ET_TARGET_DET
mV
4375
0–5000
Target detector reading
during electrical test.