Model GFC7000E Instruction Manual

APPENDIX A-2: Setup Variables For Serial I/O, Revision

E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0

045840110 Rev B.3

232

SETUP VARIABLE

NUMERIC

UNITS

DEFAULT VALUE

VALUE RANGE

DESCRIPTION

USER_UNITS

— PPM

0

PPB,

PPM,

3, 8

,

UGM, MGM

3, 8

,

Concentration units for user

interface.

DIL_FACTOR

1

0.1–1000

Dilution factor. Used only if is
dilution enabled with

FACTORY_OPT variable.

DARK_CAL_DURATION

Seconds 180,

60

4

10–600

Duration of dark cal. First
two-thirds is stabilization
period; final third is measure

period.

DARK_MEAS_MV

mV

0

-1000–1000

Dark offset for measure
reading.

DARK_REF_MV

mV

0

-1000–1000

Dark offset for reference
reading.

LIN_TARGET_CONC1

Conc

300

1–10000

Target concentration during
linearity adjustment for range

1.

LIN_NORM_CONC1

PPM

300

0.01–10000

Target concentration during
linearity adjustment
normalized for T/P for range

1.

LIN_RATIO1

— 1

0.01–100

Measure/reference

ratio

measured during linearity

adjustment for range 1.

LIN_CORRECT1

1

0.001–999.999

Linearity correction factor for
range 1.

LIN_TARGET_CONC2

Conc

300

1–10000

Target concentration during
linearity adjustment for range

2.

LIN_NORM_CONC2

PPM

300

0.01–10000

Target concentration during
linearity adjustment
normalized for T/P for range

2.

LIN_RATIO2

— 1

0.01–100

Measure/reference

ratio

measured during linearity

adjustment for range 2.

LIN_CORRECT2

1

0.001–999.999

Linearity correction factor for
range 2.

LIN_TARGET_CONC

Conc

300

1–10000

Target concentration during
linearity adjustment.

LIN_NORM_CONC

PPM

300

0.01–10000

Target concentration during
linearity adjustment

normalized for T/P.

LIN_RATIO

— 1

0.01–100

Measure/reference

ratio

measured during linearity

adjustment.

LIN_CORRECT

1

0.001–999.999

Linearity correction factor.

CO2_COMP_ENABLE

OFF

ON, OFF

ON enables CO

2

compensation; OFF disables

it.

CO2_COMP_CONC

% 0

0–20

CO

2

concentration to

compensate for.

CO_CONST1

— 700,

500

4

, 40000

3, 8

100–50000

CO

calculation

constant.

CO_CONST2

— 0.13,

0.198

8

1.448

4

,1.458

12

, 0.187

3

0–10 CO

calculation

constant.

ET_MEAS_GAIN

1

0.0001–9.9999

Electrical test gain factor for
measure reading.

ET_REF_GAIN

1

0.0001–9.9999

Electrical test gain factor for
reference reading.

ET_TARGET_DET

mV

4375

0–5000

Target detector reading
during electrical test.