Table 6-3: List of Warning Messages

MESSAGE MEANING

ANALOG CAL WARNING

The instruments A/D circuitry or one of its analog outputs is not calibrated.

BENCH TEMP WARNING

The Temperature of the optical bench is outside the specified limits.

BOX TEMP WARNING

Remote span calibration failed while the dynamic span feature was set to
turned on

CANNOT DYN SPAN

Remote zero calibration failed while the dynamic zero feature was set to
turned on

CANNOT DYN ZERO

Configuration was reset to factory defaults or was erased.

CONC ALRM1 WARNING

Concentration alarm 1 is enabled and the measured CO

2

level is ≥ the set

point.

CONC ALRM2 WARNING

Concentration alarm 2 is enabled and the measured CO

2

level is ≥ the set

point.

CONFIG INITIALIZED

Configuration storage was reset to factory configuration or erased.

DATA INITIALIZED

iDAS data storage was erased.

FRONT PANEL WARN

Firmware is unable to communicate with the front panel.

PHOTO TEMP WARNING

The temperature of the IR photometer is outside the specified limits.

REAR BOARD NOT DET

The CPU is unable to communicate with the motherboard.

RELAY BOARD WARN

The firmware is unable to communicate with the relay board.

SAMPLE FLOW WARN

The flow rate of the sample gas is outside the specified limits.

SAMPLE PRESS WARN

Sample gas pressure outside of

operational parameters.

SAMPLE TEMP WARN

The temperature of the sample gas is outside the specified limits.

SOURCE WARNING

The IR source may be faulty.

SYNC WARNING

The instrument is not properly tracking the rotation of the Gas Filter

Correlation wheel.

SYSTEM RESET

The computer was rebooted.

WHEEL TEMP WARNING

The Gas Filter Correlation wheel temperature is outside the specified limits.