MAINTENANCE SCHEDULE & PROCEDURES

04584 Rev A1

146

Table 9-1: GFC7000E Maintenance Schedule

Date Performed

Item Action Freq

Cal

Check

Req’d.

Manual

Section

Particulate

Filter

Replace

Weekly or as

needed

Yes

Verify Test

Functions

Record and

analyze

Weekly or

after any

Maintenance

or Repair

No

Pump

Diaphragm

Replace

Every 2

years

Yes

Perform

Flow Check

Check Flow

Every 6

Months

No

Perform

Leak Check

Verify Leak

Tight

Annually or

after any

Maintenance

or Repair

Yes

Pneumatic

lines

Examine

and clean

As needed

Yes if

cleaned

Cleaning Clean As

needed

Only if

cover

remv’d