Model GFC7000E Instruction Manual

Calibration Procedures

04584 Rev A1

139

7.6. Automatic Zero/Span Cal/Check (AutoCal)

The AutoCal system allows unattended periodic operation of the ZERO/SPAN valve options by
using the MGFC7000E’s internal time of day clock. AutoCal operates by executing SEQUENCES
programmed by the user to initiate the various calibration modes of the analyzer and open and

close valves appropriately. It is possible to program and run up to 3 separate sequences (SEQ1,
SEQ2 and SEQ3). Each sequence can operate in one of 3 Modes, or be disabled.

Table 7-1: AUTOCAL Modes

MODE NAME

ACTION

DISABLED

Disables the Sequence

ZERO

Causes the Sequence to perform a zero calibration/check

ZERO-SPAN

Causes the Sequence to perform a zero and span
concentration calibration/check

SPAN

Causes the Sequence to perform a span concentration
calibration/check

For each mode there are seven parameters that control operational details of the SEQUENCE.

They are:

Table 7-2: AutoCal ATTRIBUTE Setup Parameters

ATTRIBUTE NAME

ACTION

TIMER ENABLED

Turns on the Sequence timer

STARTING DATE

Sequence will operate after Starting Date

STARTING TIME

Time of day sequence will run

DELTA DAYS

Number of days to skip between each Seq. execution

DELTA TIME

Number of hours later each “Delta Days” Seq is to be run

DURATION

Number of minutes the sequence operates

CALIBRATE

Enable to do a calibration – Disable to do a cal check only