beautypg.com

Dc power supply, C bus – Teledyne GFC-7000E - Trace CO2 Analyzer User Manual

Page 211

background image

Model GFC7000E Instruction Manual

TROUBLESHOOTING & REPAIR PROCEDURES

04584 Rev A1

200

11.4.2. DC Power Supply

If you have determined that the analyzer’s AC mains power is working, but the unit is still not

operating properly, there may be a problem with one of the instrument’s switching power
supplies. The supplies can have two faults, namely no DC output, and noisy output.

To assist tracing DC Power Supply problems, the wiring used to connect the various printed circuit
assemblies and DC Powered components and the associated test points on the relay board follow

a standard color-coding scheme as defined in Table.

Table 11-6: DC Power Test Point and Wiring Color Codes

NAME

TEST POINT#

TP AND WIRE COLOR

Dgnd

1

Black

+5V

2

Red

Agnd

3

Green

+15V

4

Blue

-15V

5

Yellow

+12V

6

Purple

+12R

7

Orange

A voltmeter should be used to verify that the DC voltages are correct per the values in Table

below, and an oscilloscope, in AC mode, with band limiting turned on, can be used to evaluate if
the supplies are producing excessive noise (> 100 mV p-p).

Table 11-7: DC Power Supply Acceptable Levels

CHECK RELAY BOARD TEST POINTS

FROM TEST

POINT

TO TEST POINT

POWER

SUPPLY

ASSY

VOLTAG

E

NAME # NAME #

MIN V

MAX V

PS1 +5 Dgnd 1 +5 2 4.8 5.25
PS1 +15 Agnd 3 +15 4 13.5 16V
PS1 -15 Agnd 3 -15V 5 -14V

-16V

PS1 Agnd Agnd 3 Dgnd 1 -0.05 0.05
PS1 Chassis Dgnd 1 Chassis N/A -0.05 0.05
PS2 +12

+12V

Ret

6 +12V 7 11.75

12.5

PS2 Dgnd

+12V

Ret

6 Dgnd 1 -0.05 0.05

11.4.3. I

2

C Bus

Operation of the I

2

C buss can be verified by observing the behavior of D1 on the Relay Board in

conjunction with the performance of the front panel display. Assuming that the DC power

supplies are operating properly and the wiring from the Motherboard to the Keyboard, and the
wiring from the keyboard to the Relay board, is intact, the I

2