TROUBLESHOOTING & REPAIR PROCEDURES

04584 Rev A1

191

D1 – M/R Sensor Status

D2 – Segment Sensor Status

JP4 Connector to Opto-Pickup

Board

Figure 11-4: Sync/Demod Board Status LED Locations

11.1.4.3. Relay Board Status LED’s

There are eight LED’s located on the Relay Board. The most important of which is D1, which

indicates the health of the I

2

C buss. If D1 is blinking the other faults following LED’s can be used

in conjunction with DIAG menu signal I/O to identify hardware failures of the relays and switches
on the relay (see Section 6.9.2 and Appendix D).

Table 11-4:

I

2

C

Status LED Failure Indications

LED FUNCTION

FAULT

STATUS

INDICATED

FAILURE(S)

D1

(Red)

I2C buss

Health

(Watchdog

Circuit)

Continuously ON

or

Continuously OFF

Failed/Halted CPU

Faulty Mother Board, Keyboard or Relay

Board

Faulty Connectors/Wiring between Mother
Board, Keyboard or Relay Board

Failed/Faulty +5 VDC Power Supply (PS1)

STATUS LED’s

DC VOLTAGE TEST

POINTS

RELAY PCA

PN 04135

Figure 11-5: Relay Board Status LEDs