Model GFC7000E Instruction Manual

Optional Hardware and Software

04584 Rev A1

44

When in maintenance mode the instrument ignores all commands received via the COMM ports
that alter the operation state of the instrument This includes all calibration commands, diagnostic

menu commands and the reset instrument command. the instrument continues to measure
concentration and send data when requested.

This option is of particular use for instruments connected to multidrop or Hessen protocol

networks.

5.10. Second Language Switch

API’s instruments can be equipped with switch that activates an alternate set of display message

in a language other than the instruments default language. When present, the switch accessed by
opening the hinged front panel and is located on the rearward facing side of the display/keyboard

driver PCA; on the right side.

To activate this feature, the instrument must also have a specially programmed Disk on Chip
containing the second language.