beautypg.com

Other performance problems, Inability to span – no span key, Inability to zero – no zero key – Teledyne GFC-7000E - Trace CO2 Analyzer User Manual

Page 207

background image

Model GFC7000E Instruction Manual

TROUBLESHOOTING & REPAIR PROCEDURES

04584 Rev A1

196

4. Confirm the sample pressure, wheel temperature, bench temperature, and sample flow

readings are correct and have steady readings.

5. Disconnect the exhaust line from the optical bench near the rear of the instrument and plug

this line into the SAMPLE inlet creating a pneumatic loop. The CO

2

concentration (either zero

or span) now must be constant. If readings become quiet, the problem is in the external

pneumatics supplies for sample gas, span gas or zero air.

6. If pressurized span gas is being used with a zero/span valve option, make sure that the

venting is adequate (See Section 3.1.2 and 5.4)

7. If it is the zero point that is non-repeatable, and if that non-repeatability seems to only occur

at a certain time of day, such as when worker occupancy is highest or lowest, make sure the
flow of purge gas to the GFC wheel housing has not stopped (see Sections 10.2.2 and 11.1.7.

for more information).

11.2.3. Inability to Span – No SPAN Key

1. Confirm that the carbon dioxide span gas source is accurate; this can be done by switching

between two span-gas tanks. If the CO

2

concentration is different, there is a problem with

one of the tanks.

2. Check for leaks in the pneumatic systems as described in Section 9.3.3.

3. Make sure that the expected span gas concentration entered into the instrument during

calibration is the correct span gas concentration and not too different from expected span
value. This can be viewed via the RNG Menu (see Section 6.7).

4. Check to make sure that there is no ambient air or zero air leaking into span gas line.

11.2.4. Inability to Zero – No ZERO Key

1. Confirm that there is a good source of zero air. Dilute a tank of span gas with the same

amount of zero air from two different sources. If the CO

2

Concentration of the two

measurements is different, there is a problem with one of the sources of zero air.

2. Check for leaks in the pneumatic systems as described in Section 9.3.3.

3. The Internal zero air scrubber may need maintenance. This device is only present if the

analyzer has had zero/span valve options 51 or 53 installed this.

4. Check to make sure that there is no ambient air leaking into zero air line.

11.3. Other Performance Problems

Dynamic problems (i.e. problems which only manifest themselves when the analyzer is monitoring

sample gas) can be the most difficult and time consuming to isolate and resolve. The following
section provides an itemized list of the most common dynamic problems with recommended

troubleshooting checks and corrective actions.