Model GFC7000E Instruction Manual

Operating Instructions

04584 Rev A1

65

To access and navigate the VARS menu, use the following key sequence.

SETUP X.X

CLOCK_ADJ=0 Sec/Day

+ 0 0

ENTR EXIT

SETUP X.X

7) CLOCK_ADJ=0 Sec/Day

PREV

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

4 ) DYN_ZERO=ON

PREV

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

DYN_ZERO=ON


ON

ENTR EXIT

Toggle this keys to change setting

SETUP X.X

5) DYN_SPAN=ON

PREV

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

DYN_SPAN=ON


ON

ENTR EXIT

Toggle this keys to change setting

Toggle tese keys to change setting

SETUP X.X

CONC_PRECUISION : 3

AUTO 0 1 2 3 4 ENTR EXIT

SETUP X.X

6) CONC_PRECUISION : 3

PREV

NEXT

JUMP

EDIT

PRNT EXIT

Toggle these keys to change setting

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS

DIAG ALRM

EXIT

SETUP X.X

0 ) DAS_HOLD_OFF=

15.0 Minutes

NEXT

JUMP

EDIT

PRNT EXIT

SETUP X.X

DAS_HOLD_OFF=15.0 Minutes

1 5 .0

ENTR EXIT

EXIT ignores the new setting.

ENTR accepts the new setting.

Toggle this keys to change setting

SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP