Setting the hessen protocol response mode, Hessen protocol gas id, Text – Teledyne GFC-7000E - Trace CO2 Analyzer User Manual
Page 137
Model GFC7000E Instruction Manual
Operating Instructions
04584 Rev A1
126
NOTE
While Hessen Protocol Mode can be activated independently for COM1 and COM2, The
TYPE selection affects both Ports.
6.13.4.5. Setting The Hessen Protocol Response Mode
The Teledyne Instruments implementation of Hessen Protocol allows the user to choose one of
several different modes of response for the analyzer.
Table 6-30: Teledyne Instruments Hessen Protocol Response Modes
MODE ID
MODE DESCRIPTION
CMD
This is the Default Setting. Reponses from the instrument are encoded as the traditional
command format. Style and format of responses depend on exact coding of the initiating
command.
BCC
Responses from the instrument are always delimited with
response,
(checksum), regardless of the command encoding.
TEXT
Responses from the instrument are always delimited with
of the string, regardless of the command encoding.
To Select a Hessen response mode, press:
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG ALRM
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SETUP X.X
HESSEN RESPONSE MODE :CMD
SET> EDIT EXIT SETUP X.X COMMUNICATIONS MENU ID HESN COM1 COM2 EXIT SETUP X.X HESSEN VARIATION: TYPE 1 SET> EDIT EXIT SETUP X.X HESSEN RESPONSE MODE :CMD BCC TEXT EDIT ENTR EXIT Press to change response mode. ENTR key accepts the new settings EXIT key ignores the new settings 6.13.4.6. Hessen Protocol Gas ID The Model GFC7000E Analyzer is a single gas instrument that measures CO 2 . As such it’s default gas ID has already been set to 310. There is no need to change this setting.