Model GFC7000E Instruction Manual

APPENDIX A-2: Setup Variables For Serial I/O, Revision

E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0

045840110 Rev B.3

237

SETUP VARIABLE

NUMERIC

UNITS

DEFAULT VALUE

VALUE RANGE

DESCRIPTION

CLOCK_FORMAT

“TIME=%H:%M:%S”

Any character in
the allowed
character set. Up
to 100 characters

long.

Time-of-day clock format
flags. Enclose value in double
quotes (") when setting from

the RS-232 interface.
“%a” = Abbreviated weekday

name.
“%b” = Abbreviated month

name.
“%d” = Day of month as

decimal number (01 – 31).
“%H” = Hour in 24-hour

format (00 – 23).
“%I” = Hour in 12-hour

format (01 – 12).
“%j” = Day of year as

decimal number (001 – 366).
“%m” = Month as decimal

number (01 – 12).
“%M” = Minute as decimal

number (00 – 59).
“%p” = A.M./P.M. indicator

for 12-hour clock.
“%S” = Second as decimal

number (00 – 59).
“%w” = Weekday as decimal

number (0 – 6; Sunday is 0).
“%y” = Year without century,

as decimal number (00 – 99).
“%Y” = Year with century, as

decimal number.
“%%” = Percent sign.

ALARM_TRIGGER

3,4

Cycles 10

1–100

Concentration

alarm

trigger

sensitivity adjustment.

REF_SDEV_LIMIT

mV

50

0.1–500

Reference detector standard
deviation must be below this
limit to switch out of startup

mode.

0

Enclose value in double quotes (") when setting from the RS-232 interface.

1

Multi-range modes.

2

Hessen protocol.

3

M300EH.

4

GFC7000E.

5

M300ES.

6

Fixed range special.

7

iCHIP option.

8

M300EM.

9

FC7000E.

10

M306E.

11

Must power-cycle instrument for these options to take effect.

12

GFC7000EU.