6.8. SETUP – VARS: Using the Internal Variables

The MGFC7000E has several-user adjustable software variables, which define certain operational
parameters. Usually, these variables are automatically set by the instrument’s firmware, but can

be manually re-defined using the VARS menu. Table 6-6 lists all variables that are available within
the 818 password protected level.

Table 6-6: Variable Names (VARS) Revision B.3

No. Variable

Description

Allowed

Values

0 DAS_HOLD_OFF

Changes the internal data acquisition system (iDAS)
hold-off time, which is the duration when data are
not stored in the iDAS because the software

considers the data to be questionable. That is the
case during warm-up or just after the instrument

returns from one of its calibration modes to SAMPLE
mode. DAS_HOLD_OFF can be disabled entirely in

each iDAS channel.

Can be between

0.5 and 20

minutes

Default=15 min.

1 CONC_PRECISION

Allows the user to set the number of significant digits
to the right of the decimal point display of

concentration and stability values.

AUTO, 1, 2, 3, 4

Default=AUTO

2 DYN_ZERO

Dynamic zero automatically adjusts offset and slope

of the CO

2

response when performing a zero point

calibration during an AutoCal (Chapter 7).

ON/OFF

3 DYN_SPAN

Dynamic span automatically adjusts slope and slope

of the CO

2

response when performing a zero point

calibration during an AutoCal (Chapter 7).

Note that the DYN_ZERO and DYN_SPAN features
are not allowed for applications requiring EPA

equivalency.

ON/OFF

4 CLOCK_ADJ

Adjusts the speed of the analyzer’s clock. Choose the
+ sign if the clock is too slow, choose the - sign if
the clock is too fast.

-60 to +60 s/day