beautypg.com

Teledyne GFC-7000E - Trace CO2 Analyzer User Manual

Page 251

background image

Model 360E Instruction Manual

APPENDIX A-4: MGFC7000E Signal I/O Definitions, Revision E.0

045840110 Rev B.3

240

APPENDIX A-4: MGFC7000E Signal I/O Definitions, Revision E.0
Table A-4: GFC7000E Signal I/O Definitions, Revision E.0

SIGNAL NAME

BIT OR CHANNEL

NUMBER

DESCRIPTION

SYNC_OK

0

1 = sync. OK
0 = sync. error

1–7

Spare

Internal outputs, U8, J108, pins 18 = bits 07, default I/O

ELEC_TEST

0

1 = electrical test on
0 = off

DARK_CAL

1

1 = dark calibration on
0 = off

2–5

Spare

I2C_RESET

6

1 = reset I2C peripherals
0 = normal

I2C_DRV_RST

7

0 = hardware reset 8584 chip
1 = normal

Control inputs, U11, J1004, pins 1–6 = bits 0–5, default I/O address 321 hex

EXT_ZERO_CAL

0

0 = go into zero calibration
1 = exit zero calibration

EXT_SPAN_CAL

1

0 = go into span calibration
1 = exit span calibration

REMOTE_RANGE_HI

2

0 = remote select high range
1 = default range

3–5

Spare

6–7

Always

1

Control inputs, U14, J1006, pins 16 = bits 05, default I/O

0–5

Spare

6–7

Always

1

Control outputs, U17, J1008, pins 18 = bits 07, default I/

0–7

Spare

Control outputs, U21, J1008, pins 912 = bits 03, default I/

0–3

Spare

Alarm outputs, U21, J1009, pins 112 = bits 47, default I/O

ST_SYSTEM_OK2

4

1 = system OK
0 = any alarm condition or in diagnostics mo

ST_CONC_ALARM_1

2 + 8

5

1 = conc. limit 1 exceeded
0 = conc. OK

ST_CONC_ALARM_2

2 + 8

6

1 = conc. limit 2 exceeded
0 = conc. OK

7

Spare

A status outputs, U24, J1017, pins 18 = bits 07, default I/

ST_SYSTEM_OK

0

0 = system OK
1 = any alarm condition

ST_CONC_VALID

1

0 = conc. valid
1 = hold off or other conditions

ST_HIGH_RANGE

2

0 = high auto-range in use
1 = low auto-range

ST_ZERO_CAL

3

0 = in zero calibration
1 = not in zero