Model GFC7000E Instruction Manual

THEORY OF OPERATION

04584 Rev A1

157

calibration SLOPE and OFFSET values to produce the CO

2

concentration which is then normalized

for changes in sample pressure.

IR shining through both cells is

effected equally by interfering gas

in the Sample Chamber

M/R

is Shifted

Figure 10-5: Effects of Interfering Gas on CO2 MEAS & CO2 REF

If an interfering gas, such as H

2

O vapor is introduced into the sample chamber, the spectrum of

the IR beam is changed in a way that is identical for both the reference and the measurement

cells, but without changing the ratio between the peak heights of CO

2

MEAS and CO

2

REF. In

effect, the difference between the peak heights remains the same.

Thus, the difference in the peak heights and the resulting M/R ratio is only due to CO

2

and not to

interfering gases. In this way, Gas filter correlation rejects the effects of interfering gases and so

that the analyzer responds only to the presence of CO

2

.

To improve the signal-to-noise performance of the IR photo-detector, the GFC wheel also

incorporates an optical mask that chops the IR beam into alternating pulses of light and dark at
six times the frequency of the measure/reference signal. This limits the detection bandwidth
helping to reject interfering signals from outside this bandwidth improving the signal to noise

ration.

The IR Signal as the Photo-Detector

sees it after being chopped by the GFC

Wheel Screen

CO REF

CO MEAS

Figure 10-6: Chopped IR Signal