Model GFC7000E Instruction Manual

Operating Instructions

04584 Rev A1

102

These default Data Channels can be used as they are, or they can be customized from the front
panel to fit a specific application. They can also be deleted to make room for custom user-

programmed Data Channels.

Appendix A-5 lists the firmware-specific iDAS configuration in plain-text format. This text file can

either be loaded into APICOM and then modified and uploaded to the instrument or can be copied
and pasted into a terminal program to be sent to the analyzer.

NOTE

Sending an iDAS configuration to the analyzer through its COM ports will replace the

existing configuration and will delete all stored data. Back up any existing data and the

iDAS configuration before uploading new settings.

These default Data Channels can be used as they are, or they can be customized from the front
panel to fit a specific application. They can also be deleted to make room for custom user-

programmed Data Channels.

Appendix A-5 lists the firmware-specific iDAS configuration in plain-text format. This text file can

either be loaded into APICOM and then modified and uploaded to the instrument or can be copied
and pasted into a terminal program to be sent to the analyzer.

NOTE

Sending an iDAS configuration to the analyzer through its COM ports will replace the

existing configuration and will delete all stored data. Back up any existing data and the

iDAS configuration before uploading new settings.