Model GFC7000E Instruction Manual

Operating Instructions

04584 Rev A1

107

To configure a specific data parameter, press:

FROM THE EDIT DATA PARAMETER MENU

(see previous section)

SETUP X.X

0) PARAM=CONC1, MODE=AVG


PREV NEXT INS DEL

EDIT

EXIT

SETUP X.X

PARAMETERS:CONC!

SET> EDIT

EXIT

SETUP X.X

SAMPLE MODE:AVG

EDIT

EXIT

SETUP X.X

PARAMETERS: PMTDET

PREV NEXT

ENTR

EXIT

If more than on parameter is active for

this channel, these cycle through list of

existing Parameters.

ENTR accepts the new

setting and returns to the

previous menu.

EXIT ignores the new setting

and returns to the previous

SETUP X.X

SAMPLE MODE: INST

INST AVG MIN MAX

EXIT

Press the key for the desired mode

SETUP X.X

PRECISION: 1

EDIT

EXIT

SETUP X.X

PRECISION: 1

1

EXIT

Set for 0-4

SETUP X.X

STORE NUM. SAMPLES: OFF

EDIT

EXIT

SETUP X.X

STORE NUM. SAMPLES: OFF

OFF

ENTR EXIT

Turn ON or OFF

Returns to

previous

Functions

6.12.2.5. Sample Period and Report Period

The iDAS defines two principal time periods by which sample readings are taken and permanently
recorded:

SAMPLE PERIOD: Determines how often iDAS temporarily records a sample reading of

the parameter in volatile memory. The SAMPLE PERIOD is set to one minute by default

and generally cannot be accessed from the standard iDAS front panel menu, but is
available via the instruments communication ports by using APICOM or the analyzer’s
standard serial data protocol.

SAMPLE PERIOD is only used when the iDAS parameter’s sample mode is set for AVG,
MIN
or MAX.