Model GFC7000E Instruction Manual

Getting Started

04584 Rev A1

23

The analyzer should automatically switch to SAMPLE mode after completing the boot-up
sequence and start monitoring CO

2

gas. Warm-Up

3.2.2. Warm Up

The MGFC7000E requires about 30 minutes warm-up time before reliable CO

2

measurements can

be taken. During that time, various portions of the instrument’s front panel will behave as follows.

See Figure 3-10 for locations.

Table 3-5: Front Panel Display During System Warm-Up

Name Color Behavior

Significance

Concentration

Field

N/A Displays

current,

compensated CO

2

Concentration

N/A

Mode Field

N/A

Displays blinking

“SAMPLE”

Instrument is in sample mode but is still in the
process of warming up.

STATUS LED’s

Sample Green

On

Unit is operating in sample mode, front panel

display is being updated.
Flashes On/Off when adaptive filter is active

Cal

Yellow

Off

The instrument’s calibration is not enabled.

Fault

Red

Blinking

The analyzer is warming up and hence out of
specification for a fault-free reading. various warning

messages will appear.

MODE FIELD

KEY DEFINITIONS

MESSAGE FIELD

CONCENTRATION FIELD STATUS LED’s

KEYBOARD

ON / OFF SWITCH

FASTENER

FASTENER

POWER

FAULT

CAL

SAMPLE

GAS FILTER CORRELATION CO

2

ANALYZER- MODEL GFC7000E

SAMPLE A

RANGE = 500.0 PPM

CO2 = 400.0

TST>

CAL

SETUP

LOCKING SCREW

Figure 3-10: Front Panel Layout