Model GFC7000E Instruction Manual

TROUBLESHOOTING & REPAIR PROCEDURES

04584 Rev A1

190

because it may be possible to recover operation of the analyzer. If after 30 – 60 seconds neither
the DS5 is flashing or no characters have been written to the front panel display then the CPU is

bad and must be replaced.

Mother Board

P/N 04069

CPU Status LED

Figure 11-3: CPU Status Indicator

11.1.4.2. Sync Demodulator Status LED’s

Two LED’s located on the Sync/Demod Board and are there to make it obvious that the GFC

Wheel is spinning and the synchronization signals are present:

Table 11-3: Sync/Demod Board Status Failure Indications

LED

Function

Fault Status

Indicated Failure(s)

D1 M/R

Sensor

Status

LED is stuck

ON or OFF

GFC Wheel is not turning
M/R Sensor on Opto-Pickup Board failed

Sync/Demod Board failed

JP 4 Connector/Wiring faulty
Failed/Faulty +5 VDC Power Supply (PS1)

D2 Segment

Sensor Status

LED is stuck

ON or OFF

GFC Wheel is not turning
Segment Sensor on Opto-Pickup Board failed

Sync/Demod Board failed

JP 4 Connector/Wiring faulty

Failed/Faulty +5 VDC Power Supply (PS1)