Model GFC7000E Instruction Manual

Getting Started

04584 Rev A1

13

Table 3-1: GFC7000E Analog Output Pin Outs

Pin

Analog Output

VDC Signal

mADC Signal

1

V Out

I Out +

2

A1

Ground

I Out -

3

V Out

I Out +

4

A2

Ground

I Out -

5

V Out

I Out +

6

A3

Ground

I Out -

7

V Out

Not Available

8

A4 (Spare)

Ground Not

Available

• The default analog output voltage setting of the GFC7000E CO

2

Analyzer is 0 – 5 VDC

with a range of 0 – 500 ppm.

• TO change these settings, see Sections 6.9.4 and 6.7 respectively.

10. If you wish utilize the analyzer’s status outputs to interface with a device that accepts logic-

level digital inputs, such as programmable logic controllers (PLC’s) they are accessed via a 12-

pin connector on the analyzer’s rear panel labeled STATUS.

STATUS

1 2 3 4 5 6 7 8 D

+

NOTE

Most PLC’s have internal provisions for limiting the current that the input will draw from

an external device. When connecting to a unit that does not have this feature, an

external dropping resistor must be used to limit the current through the transistor

output to less than 50mA. At 50mA, the transistor will drop approximately 1.2V from its

collector to emitter.