Model GFC7000E Instruction Manual

Calibration Procedures

04584 Rev A1

131

MODEL 700

Gas Dilution

Calibrator

VENT

Calibrated

CO

2

Gas

Source of

SAMPLE Gas

Removed

during

Calibration

MODEL

GFC7000E

Sample

IZS

Exhaust

Vent Span

Pressure Span

Purge In

MODEL 701

Zero Air Generator

Indicating

soda-lime

Figure 7-2: Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator

STEP TWO: Set the expected CO

2

Span Gas concentration:

This sequence causes the

analyzer to prompt for the

expected CO

x

span

concentration.

M-P CAL

SO2 SPAN CONC: 400.000 Conc

0 0 0

4 5 .0

ENTR

EXIT

M-P CAL RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > ZERO

CONC

EXIT

SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX

< TST TST >

CAL

SETUP

The SO

2

span

concentration values

automatically default to

400.0 Conc.

To change this value to

the actual concentration of

the span gas, enter the

number by pressing the

key under each digit until

the expected value

appears.

EXIT ignores the new setting

and returns to the previous

display.

ENTR accepts the new setting

and returns to the

previous display..

NOTE

For this Initial Calibration it is important to independently verify the PRECISE CO

2

Concentration Value of the SPAN gas.

If the source of the Span Gas is from a Calibrated Bottle, use the exact concentration

value printed on the bottle.