Model GFC7000E Instruction Manual

Calibration Procedures

04584 Rev A1

138

M-P CAL

STABIL=XXX.X PPM

CO2 =XXX.X

< TST

TST

>

CAL

CALZ

CALS SETUP

ACTION:

Allow zero gas to enter the sample port at the

rear of the instrument.

Wait until STABIL

falls below 1.0 ppb.

This may take several

minutes.

SAMPLE* RANGE = 500.000 PPM CO2 =XXX.X

TST>

CAL

CALZ CALS SETUP

SAMPLE

STABIL=XXX.X PPM

CO2 =XXX.X

< TST TST >

CAL

CALZ CALS SETUP

Set the Display to show the

STABIL test function.

This function calculates the

stability of the CO

x

measurement

Continue Calibration as per

Standard Procedure

ZERO CAL M

RANGE = 500.000 PPB SO2 =X.XXX

< TST TST >

ZERO

SPAN

CONC

EXIT

ANALYZER ENTERS

ZERO CAL MODE

SAMPLE

RANGE TO CAL: LOW

LOW

HIGH

ENTR SETUP

SAMPLE

RANGE TO CAL: HIGH

LOW HIGH

ENTR

SETUP

Once this selection is made, the calibration procedure continues as previously described in Section

7.2. The other range may be calibrated by starting over from the main SAMPLE display.

7.5.2. Use of Zero/Span Valves with Remote Contact Closure

Contact closures for controlling calibration are located on the rear panel CONTROL IN connector.

Instructions for setup and use of these contacts are found in Section 6.11.1.2. When the contacts
are closed for at least 5 seconds, the instrument switches into zero or span mode. The remote

calibration contact closures may be activated in any order. It is recommended that contact
closures remain closed for at least 10 minutes to establish a reliable reading.

The instrument will stay in the selected mode for as long as the contacts remain closed. If
calibration is enabled, the MGFC7000E will re-calibrate when the contact is opened, then go into

SAMPLE mode. If calibration is disabled, the instrument will return to SAMPLE mode, leaving the
calibration unchanged.