Teledyne GFC-7000E - Trace CO2 Analyzer User Manual

Page 81

background image

Model GFC7000E Instruction Manual

Operating Instructions

04584 Rev A1

70

The following DC current output limits apply to the current loop modules:

Table 6-10: Analog Output Current Loop Range

RANGE

MINIMUM OUTPUT

MAXIMUM OUTPUT

0-20 mA

0 mA

20 mA

These are the physical limits of the current loop modules, typical

applications use 2-20 or 4-20 mA for the lower and upper limits.
Please specify desired range when ordering this option.
The default offset for all ranges is 0 mA.

Pin assignments for the output connector at the rear panel of the instrument are shown in Table

6-11.

ANALOG OUT

A1 A2 A3 A4
+ - + - + - + -

Table 6-11: Analog Output Pin Assignments

PIN ANALOG

OUTPUT

VOLTAGE

SIGNAL

CURRENT

SIGNAL

1

V Out

I Out +

2

A1

Ground

I Out -

3

V Out

I Out +

4

A2

Ground

I Out -

5

V Out

I Out +

6

A4

Ground

I Out -

7 & 8

A3

Not Used

Not Used

See Figure 3-2 for a the location of the analog output connector on the instruments rear panel.