Analog output step test – Teledyne GFC-7000E - Trace CO2 Analyzer User Manual

Page 79

background image

Model GFC7000E Instruction Manual

Operating Instructions

04584 Rev A1

68

To enter the signal I/O test mode, press:

SAMPLE RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

EXAMPLE

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

DIAG

SIGNAL I / O

PREV NEXT

JUMP

ENTR

EXIT

DIAG I / O

Test Signals Displayed Here

PREV NEXT

JUMP

PRNT EXIT

Use the JUMP key to

go directly to a

specific signal

See Appendix A-4 for

a complete list of

available SIGNALS

DIAG I / O

JUMP TO: 12

1 2

ENTR

EXIT

EXAMPLE:

Enter 12 to Jump to

12) ST_CONC_VALID

DIAG I / O

ST_CONC_VALID = ON

PREV NEXT

JUMP

ON

PRNT

EXIT

Exit to return

to the

DIAG menu

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

ALRM

EXIT

Use the NEXT & PREV

keys to move between

signal types.

Pressing the PRNT key will send a formatted printout to the serial port and can be

captured with a computer or other output device.

EXIT returns

to the main

SAMPLE display

6.9.3. Analog Output Step Test

This test can be used to check the accuracy and proper operation of the analog outputs. The test
forces all four analog output channels to produce signals ranging from 0% to 100% of the full

scale range in 20% increments. This test is useful to verify the operation of the data
logging/recording devices attached to the analyzer.

To begin the Analog Output Step Test press:



SAMPLE RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

DIAG

SIGNAL I / O

NEXT

ENTR

EXIT

Exit-Exit

returns to the

DIAG menu

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

ALRM

EXIT

DIAG

ANALOG OUTPUT

PREV

NEXT

ENTR

EXIT

DIAG AOUT

ANALOG

OUTPUT

0%

EXIT

Performs

analog output

step test.

0% - 100%

DIAG AOUT

ANALOG OUTPUT

[0%]

EXIT

Pressing the key under “0%” while performing the test will

pause the test at that level. Brackets will appear around
the value: example: [20%] Pressing the same key again

will resume the test.