Model GFC7000E Instruction Manual

Operating Instructions

04584 Rev A1

61

6.7.5. Auto Range Mode (AUTO)

In AUTO range mode, the analyzer automatically switches the reporting range between two user-

defined ranges (low and high). The unit will switch from low range to high range when the CO

2

concentration exceeds 98% of the low range span. The unit will return from high range back to

low range once both the CO

2

concentration falls below 75% of the low range span.

In AUTO Range mode the instrument reports the same data in the same range on both the A1

and A2 outputs and automatically switches both outputs between ranges as described above.
Also, the RANGE test function displayed on the front panel will automatically switch to show

which range is in effect.

The high/low range status is also reported through the external, digital status bits (Section
6.13.1.1).

To set individual ranges press the following keystroke sequence.



SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SETUP X.X

RANGE MODE:

SNGL

SNGL DUAL

AUTO

ENTR EXIT

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

EXIT x 2 returns

to the main

SAMPLE display

SETUP X.X

RANGE MODE:

AUTO

SNGL

DUAL AUTO

ENTR

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET

UNIT

EXIT

SETUP X.X

HIGH RANGE:

500.0 Conc

0 0 5 0 0 .0

ENTR

EXIT

SETUP X.X

LOW RANGE:

500.0 Conc

0 0 5 0 0 .0

ENTR

EXIT

Toggle the numeral

keys to set the

LOW and HIGH

range value.

ENTR accepts the

new setting, EXIT

ignores the new

setting.

CAUTION

In AUTO range mode the LOW and HIGH ranges have separate slopes and offsets for

computing CO2 concentration.

The two ranges must be independently calibrated.

NOTE

Avoid accidentally setting the low range of the instrument with a higher span limit than

the high range. This will cause the unit to stay in the low reporting range perpetually

and defeat the function of the AUTO range mode.