Model GFC7000E Instruction Manual

Operating Instructions

04584 Rev A1

90

6.10.9.1. Ethernet Card COM2 Communication Modes and Baud Rate

The firmware on board the Ethernet card automatically sets the communication modes and baud
rate (115 200 kBaud ) for the COM2 port. Once the Ethernet option is installed and activated,

the COM2 submenu is replaced by a new submenu, INET. This submenu is used to manage and
configure the Ethernet interface with your LAN or Internet Server(s).

6.10.9.2. Configuring the Ethernet Interface Option using DHCP

The Ethernet option for you MGFC7000E uses Dynamic Host Configuration Protocol (DHCP) to
automatically configure its interface with your LAN. This requires your network servers also be

running DHCP. The analyzer will do this the first time you turn the instrument on after it has been
physically connected to your network. Once the instrument is connected and turned on it will
appear as an active device on your network without any extra set up steps or lengthy procedures.

Should you need to, the following Ethernet configuration properties are viewable via the analyzer’s
front panel.

Table 6-18: LAN/Internet Configuration Properties

PROPERTY DEFAULT

STATE

DESCRIPTION

DHCP

STATUS

On Editable

This displays whether the DHCP is turned ON or
OFF.

INSTRUMENT

IP ADDRESS

Configured

by DHCP

EDIT key

disabled

when DHCP

is ON

This string of four packets of 1 to 3 numbers each
(e.g. 192.168.76.55.) is the address of the
analyzer itself.

GATEWAY IP

ADDRESS

Configured

by DHCP

EDIT key

disabled

when DHCP

is ON

A string of numbers very similar to the Instrument

IP address (e.g. 192.168.76.1.)that is the address
of the computer used by your LAN to access the
Internet.

SUBNET

MASK

Configured

by DHCP

EDIT key

disabled

when DHCP

is ON

Also a string of four packets of 1 to 3 numbers
each (e.g. 255.255.252.0) that defines that

identifies the LAN the device is connected to.
All addressable devices and computers on a LAN
must have the same subnet mask. Any

transmissions sent devices with different assumed
to be outside of the LAN and are routed through

gateway computer onto the Internet.

TCP PORT

3000 Editable

This number defines the terminal control port by
which the instrument is addressed by terminal
emulation software, such as Internet or Teledyne

Instruments’ APICOM.

HOST NAME MGFC7000E

Editable

The name by which your analyzer will appear
when addressed from other computers on the LAN

or via the Internet. While the default setting for
all Teledyne Instruments MGFC7000E analyzers is

“mGFC7000E” the host name may be changed to
fit customer needs.

1

Do not change the setting for this property unless instructed to by Teledyne Instruments