Model GFC7000E Instruction Manual

MAINTENANCE SCHEDULE & PROCEDURES

04584 Rev A1

145

9. MAINTENANCE SCHEDULE & PROCEDURES

Predictive diagnostic functions including failure warnings and alarms built into the analyzer’s

firmware allow the user to determine when repairs are necessary without performing painstaking
preventative maintenance procedures. There are, however, a minimal number of simple

procedures that when performed regularly will ensure that the analyzer continues to operate
accurately and reliably over its the lifetime. Repairs and troubleshooting are covered in Chapter 11

of this manual.

9.1. Maintenance Schedule

Table 9-1 shows a typical maintenance schedule for the analyzer. Please note that in certain
environments (i.e. dusty, very high ambient pollutant levels) some maintenance procedures may

need to be performed more often than shown.

NOTE

A Span and Zero Calibration Check (see CAL CHECK REQ’D Column of Table 9-1) must be

performed following certain of the maintenance procedure listed below.

See Sections 7.3, 7.5 and 7.6 for instructions on performing checks.

CAUTION

Risk of electrical shock. Disconnect power before performing any of the

following operations that require entry into the interior of the analyzer.

NOTE

The operations outlined in this chapter are to be performed by qualified

maintenance personnel only.