beautypg.com

Specifications, approvals and warranty, Specifications – Teledyne GFC-7000E - Trace CO2 Analyzer User Manual

Page 16

background image

Model GFC7000E Instruction Manual

Specifications, Approvals and Warranty

04584 Rev A1

5

2. SPECIFICATIONS, APPROVALS AND

WARRANTY

2.1. Specifications

Table 2-1: Model GFC7000E Basic Unit Specifications

Min/Max Range
(Physical Analog Output)

In 1ppb increments from 50ppb to 2 000ppm, dual ranges or auto
ranging

Measurement Units

ppb, ppm, µg/m

3

, mg/m

3

, %(user selectable)

Zero Noise

< 0.1 ppm (RMS)

Span Noise

< 1% of reading (RMS)

Lower Detectable Limit1

< 0.2 ppm

1

Zero Drift (24 hours)

<0.25 ppm

1

Zero Drift (7 days)

<0.5 ppm

1

Span Drift (7 Days)

1% of reading above 50 PPM

1

Linearity

1% of full scale

Precision

0.5% of reading

Temperature Coefficient

< 0.1% of Full Scale per oC

Voltage Coefficient

< 0.05% of Full Scale per V

Lag Time

10 sec

Rise/Fall Time

95% in <60 sec

Sample Flow Rate

800cm

3

/min. ±10%

Temperature Range

5-40

o

C

Humidity Range

0 - 95% RH, non-condensing

Dimensions H x W x D

7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)

Weight, Analyzer

38 lbs. (17 kg); add 1 lbs (0.5 kg) for IZS

AC Power Rating

100 V, 50/60 Hz (3.25A);
115 V, 60 Hz (3.0 A);

220 – 240 V, 50/60 Hz (2.5 A)

Environmental

Installation category (over-voltage category) II; Pollution degree 2

Analog Outputs

Three (3) Outputs

Analog Output Ranges

0.1V, 1 V, 5 V, 10 V, 2-20 or 4-20 mA isolated current loop.

All Ranges with 5% Under/Over Range

Analog Output Resolution

1 part in 4096 of selected full-scale voltage

Status Outputs

8 Status outputs - opto-isolated; including 2 alarm outputs

Control Inputs

6 Control Inputs, 3 defined, 3 spare

Serial I/O

One (1) RS-232; One (1) RS-485
Baud Rate : 300 – 115200: Optional Ethernet Interface

Certifications

CE: EN61010-1:90 + A1:92 + A2:95, EN61326 - Class A

1

At constant temperature and voltage.