Model GFC7000E Instruction Manual

Operating Instructions

04584 Rev A1

62

6.7.6. Range Units

The MGFC7000E can display concentrations in parts per billion (10

9

mols per mol, PPB), parts per

million (10

6

mols per mol, PPM), micrograms per cubic meter (µg/m

3

, UG), milligrams per cubic

meter (mg/m

3

, MG) or percent (volume CO

2

/volume sample gas, %). Changing units affects all of

the display, analog outputs, COM port and iDAS values for all reporting ranges regardless of the
analyzer’s range mode.

NOTE

Concentrations displayed in mg/m3 and ug/m3 use 0°C, 760 mmHg for Standard

Temperature and Pressure (STP). Consult your local regulations for the STP used by

your agency.

Conversion factors from volumetric to mass units are:

CO

2

: ppb x 1.96 = µg/m3; ppm x 1.96 = mg/m3

To change the concentration units:

SETUP X.X

CONC UNITS: %

PPM PPB

UGM

MGM %

ENTER EXIT

SAMPLE RANGE = 500.00 PPB CO2=X.XXX

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET

UNIT

EXIT

SETUP X.X

CONC UNITS: PPM

PPM PPB

UGM

MGM

%

ENTER

EXIT

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

Select the preferred

concentration unit.

EXIT returns

to the main menu.

ENTR accepts

the new unit,

EXIT returns

to the SETUP

menu.

NOTE

Once the units of measurement have been changed the unit MUST be recalibrated, as

the “expected span values” previously in effect will no longer be valid. Simply entering

new expected span values without running the entire calibration routine is not

sufficient.