beautypg.com

Interface, Figure 10-15: power distribution block diagram – Teledyne GFC-7000E - Trace CO2 Analyzer User Manual

Page 185

background image

Model GFC7000E Instruction Manual

THEORY OF OPERATION

04584 Rev A1

174

CAUTION

Should the AC power circuit breaker trip, investigate and correct the
condition causing this situation before turning the analyzer back on.

AC POWER

ENTRANCE

ON/OFF

SWITCH

Cooling Fan

Mother

Board

CPU

PS 1 (+5 VDC; ±15 VDC)

PS 2 (+12 VDC)

Display

Keypad

Pressure

Sensors

IR Source

Pump

GFC

Wheel

Motor

Sync/Demod

M/R &

Segment

Sensors

Heaters

Valve

Options

KEY

AC POWER

DC POWER


RELAY

BOARD

Figure 10-15: Power Distribution Block Diagram

10.4. Interface

The analyzer has several ways to communicate the outside world, see Figure 10-16. Users can

input data and receive information directly via the Front panel keypad and display. Direct
communication with the CPU is also available by way of the analyzer’s RS232 & RS485 I/O ports

or an optional Ethernet port. The analyzer can also send and receive different kinds of information
via its external digital I/O connectors and the three analog outputs located on the rear panel.