Model GFC7000E Instruction Manual

Getting Started

04584 Rev A1

21

Some applications may require multipoint calibration checks where span gas of several different
concentrations is needed. We recommend using high-concentration, certified, calibration gas

supplied to the analyzer through a Gas Dilution Calibrator such as a Teledyne Instruments Model
700. This type of calibrator precisely mixes Span Gas and Zero Air to produce any concentration
level between 0 ppm and the concentration of the calibrated gas.

Figure 3-8 depicts the pneumatic set up in this sort of application of a Model GFC7000E CO

2

Analyzer with zero/span/shutoff valve option 50 installed (a common configuration for this type of

application).

VENT if input is pressurized

VENT

VENT

MODEL 700

Gas Dilution

Calibrator

Indicating

soda-lime

MODEL 701

Zero Air Generator

Calibrated

CO

2

Gas

Source of

SAMPLE Gas

MODEL

GFC7000E

Sample

IZS

Exhaust

Vent Span

Pressure Span

Purge In

VENT

Gas Pressure should

be regulated at

30 – 35 PSIG

Needle

valve to

control flow

Figure 3-9: Example of Pneumatic Set up for Multipoint Calibration of M360

3.2. Initial Operation

If you are unfamiliar with the MGFC7000E theory of operation, we recommend that you read
Chapter 10.

For information on navigating the analyzer’s software menus, see the menu trees described in

Appendix A.1.

NOTE

The analyzer’s cover must be installed to ensure that the temperatures of the GFC

wheel and absorption cell assemblies are properly controlled.