Model GFC7000E Instruction Manual

Operating Instructions

04584 Rev A1

55

In order to compensate for CPU clocks which run faster or slower, you can adjust a variable called
CLOCK_ADJ to speed up or slow down the clock by a fixed amount every day. To change this

variable, press:

SAMPLE RANGE = 500.000 PPM

CO2

=X.XXX

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU


CFG DAS

RNGE

PASS CLK

MORE

EXIT

3x EXIT returns

to the main SAMPLE display

SETUP X.X 0 ) DAS_HOLD_OFF=15.0 Minutes

NEXT

JUMP

EDIT PRNT EXIT

Enter sign and number of seconds per

day the clock gains (-) or loses (+).

SETUPX.X 1 ) CONC_PRECISION = 3

PREV

NEXT

JUMP

EDIT PRNT EXIT

SETUP X.X

4) CLOCK_ADJ=0 Sec/Day

PREV JUMP

EDIT

PRNT EXIT

SETUP X.X

CLOCK_ADJ:

0 Sec/Day

+ 0 0

ENTR

EXIT

SETUP X.X

4) CLOCK_ADJ=0 Sec/Day

PREV NEXT JUMP

EDIT PRNT

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS

DIAG

EXIT

Continue to press NEXT until …